Technical Data: |
Operation: |
AED Semi-Automatic; ECG Monitoring capability |
Energies: |
Standard max 200J or Power max 360J |
Waveform: |
Adaptive BTE (biphasic truncated exponential) conforming to patient chest’s impedance |
Protocols: |
Various adult shock protocols available on request |
Factory default: |
Adult Standard escalating 150, 200, 200J
Adult Power escalating 200, 250, 360J
Pediatric (Standard or Power) 50J fixed |
Charging time: |
≤9 seconds with a new and fully charged battery depleted battery will result in a longer charging time |
Analysis time: |
IEC/EN 60601-2-4 from 4 to 15 seconds |
Impedance: |
20-200 ohms |
Sensitivity: |
IEC/EN 60601-2-4 (AHADB, MITDB source), 97% |
Specificity: |
IEC/EN 60601-2-4 (AHADB, MITDB source), 99% |
Controls: |
2 buttons: ON/OFF, shock button, and 3 buttons to surf the menu. |
Indicators: |
Status LED indicator informing on device condition
Battery gauge with remaining capacity rate
Audible alerts and text display with service alarms |
Upgradeable: |
through a USB cable or memory card or by remote through AMISAVERCLOUD |
ECG Monitoring:
|
|
Operations: |
Through defibrillation pads or standard ECG electrodes attached to a separate 2-Lead patient monitoring reusable cable SAV-C0017 |
ECG size: |
Manual setting through the menu |
Heart Rate: |
30-200 bpm |
Sweep Speed: |
25 mm/sec |
Standards: |
IEC/EN 60601-2-27 less then the points 202.6.2.101; 201.12.1.101.12,13; 208.6.6.2.101 not performed for the intended use of the device, as it is not intended for environments such as operating theatres or intensive care units |
Display: |
5,7” TFT colour, 640 x 480 pixel |
Battery Options: |
|
Type: |
Li-SOCl2 Disposable, code SAV-C1032 |
Autonomy: |
250 complete rescue cycles (shocks at 200J and CPR) or 160 complete rescue cycles (shocks at 360J and CPR) or 24 hours ECG analysis for a new and fully charged battery (*) |
Shelf-Life: |
when stored in original packaging 5 years (*) |
Battery-Life: |
4 years once installed to AED, assuming one battery insertion test and daily self-test but without switching AED on (*) |
Type: |
Li-ion Accumulator, code SAV-C1033 |
Recharging time: |
2,5 hours with the charger station code SAV-C1034 (*) (recommended to charge every 4 months at least) |
Autonomy: |
200 shocks at 200J or 110 shocks at 360J or 14 hours in ECG Monitoring for a new fully charged accumulator (*) |
Battery-Life: |
2 years or 300 charging cycles (*) |
Pads Options: |
|
Type: |
Disposable, pre-gelled and self-adhesive |
Adult: |
Code SAV-C0846, for patient >8 years or >25 kg |
Pediatric: |
Code SAV-C0016, for patient <8 years or <25 kg |
Cable length: |
120 cm |
Shelf-Life: |
30 months |
Event Recording: |
|
Internal memory: |
up to 6 continuous hours of ECG and rescue events |
Optional memory: |
Removable SD card; length of storage depends on card capacity: a 2GB card records up to 100 hours |
Data recording: |
“AED1LOG“ text file with detailed self-test activity
“AEDFILES“ multimedia files with complete recorded information |
Event review: |
“Saver View Express“ data manager software |
Physical: |
|
Size: |
29,5 x 23 x 11,5 cm |
Weight: |
+/- 2,85 Kg |
Environmental: |
|
Operating temperature: |
0°C to 55°C (32°F TO 131°F) |
Storing/Shipping temperature: |
-40°C to 70°C (-40°F TO 158°F) without battery |
Humidity: |
10% to 95% relative humidity non condensing |
Sealing (IP Protection): |
IEC/EN 60529 class IP56; splash proof, dust protected |
Shock/Drop Abuse Endurance: |
IEC/EN 60601-1 clause 21; 1 meter drop, impact, force, rough handling, mobile tolerance |
Electrostatic Discharge: |
IEC/EN 61000-4-2 |
Electromagnetic Compatibility: |
IEC/EN 60601-1-2 Emission, Immunity |
Electrical Protection: |
IEC/EN 60601-1; Internally Powered, Type BF |
Directive 93/42/CEE and 2007/47/CE: |
Class IIb |
Geoloc Module: |
|
Frequency: |
GSM: 850, 900, 1800, 1900 MHz;
UMTS: 900, 2100 MHz;
GPS: 1575, 1600 MHz |
Geoloc Battery Options: |
|
Type: |
Li-SOCl2 Disposable, code SAV-C1038 |
Shelf-Life: |
when stored in original packaging 5 years (*) |
Battery-Life: |
4 years once installed to AED, assuming one battery insertion test and daily self-test but without switching AED on (*) |
Type: |
Li-ion Accumulator, code SAV-C1039 |
Recharging time: |
2,5 hours with the charger station code SAV-C1040 (*) (recommended to charge every 4 months at least) |
Battery-Life: |
2 years or 300 charging cycles (*) |
(*)Temperature at 20°C Humidity 45% non-condensing |
|